Products Low Temperature
« Back to Products

Diacell® LN2-DAC System


The Diacell® LN2-DAC System is a liquid nitrogen bath cryostat with integrated high pressure diamond anvil cell.
Simplicity and cost effectiveness are the key factors of the Diacell® LN2-DAC system. The liquid nitrogen cooled cryostat provides a variable temperature sample environment with no valves or adjustments required (temperature controller is included).
 
Designed with versatility in mind, the system comes with the Diacell® CryoDAC-Nitro integrated (fitted with type Ia diamond anvils), enabling work down to 77 K and up to 50 GPa. Pressure can be changed in the cell whilst cold, since the CryoDAC-Nitro is gas membrane driven.
Four quartz windows are standard on the cryostat, which also has a port enabling the feedthrough of the gas membrane line.  An optional fibre optic or electrical feedthrough port is also available.
Temperature below 77 K can be achieved (down to around 65 K) can be achieved with a vacuum level. In this case an optional pumping manyfold is required (not supplied).
 
Note: Gas controller is sold separately. 
A turbo pump is recommended to evacuate the vacuum shroud. A primary pump with a nitrogen cold trap can also be used but in this case, the thermal performance will not be optimal, leading to a slightly higher base temperature.
 

TECHNICAL SPECIFICATION

Cryostat Liquid nitrogen bath
Temperature Range 77 to 300 K
Cool down time Approx 15 mins
Diamond Anvil Cell Diacell® CryoDAC-Nitro
Pressure Mechanism Gas membrane
Maximum Pressure 50 GPa
DAC Top Angle Optical 52º conical
DAC Bottom Angle Optical 52º conical
Cooling Via cryostat
DAC Diameter 38 mm
DAC Height 29 mm
DAC working distance to sample 11 mm

Maximum pressure is dependent upon the culet size of the anvils.
Almax easyLab is committed to its policy of continuous improvement.
Specifications may change without notice.
easyLab and Diacell are registered trademarks of Almax easyLab Ltd.

 


PARAMETERS

EXPERIMENTAL
TECHNIQUES
AC Transport | DC Resistivity | far-IR | Fluorescence | FTIR | Hall Effect | Infrared | I-V Curves | Mossbauer | Raman | X-Ray
MAX
PRESSURE
1 GPa | 10 GPa | 3.0 GPa | 30 GPa | 50 GPa
TEMPERATURE
RANGE
Low Temperatures | Room Temperature