Diacell® Bragg-(G) LSH (Large Sight Holes)
X-ray Spectroscopy, Pmax = 100 GPa, T = 300 K, WD = 15 mm, A = 85°, 2 x 130° azimuthal openings
Diacell® Bragg Series
- The Diacell® Bragg-(G) LSH is the large side holes variation of the Bragg-(G) Plus;
- The 2 side ports on the Diacell® Bragg-(G) LSH have openings of 130° (azimuthal) by 25° (axial);
- The DAC uses conical Boehler-Almax anvils, making this cell the true benchmark for high pressure X-ray studies;
- The large angle apertures enable diffraction work with high transmission factor and very low background;
- Being a gas membrane driven means that pressure within the cell can be changed whilst the sample is mounted on the X-ray stage, saving considerable time;
- Maximum pressures of up to above 100 GPa may be obtained with the Diacell® Bragg-(G) LSH.
Log In or register to see prices.
Description
The Diacell® Bragg-(G) LSH is a variation of the Diacell® Bragg-(G) Plus, one of our most popular cells from our catalog.
This DAC has 2 wide lateral sight holes. They have an elliptical shape to enable up to 25° and 130° degrees of respectively vertical and horizontal apertures.
The Bragg-(G) LSH incorporates conical anvils mounted on tungsten carbide support plates and it is gas membrane driven.
The benefits of this type of drive include the ability to apply very small incremental loads, remote operation and the achievement of higher pressures due to absence of friction associated with bolts.
Applications
This cell is ideal for X-Ray diffraction and related work. The DAC is supplied with a 3.10 mm mounting pin that is compatible with most commercially available goniometer heads, such as those from Huber.
Specifications
| Cell Material | AISI 440C |
| Anvil Design | Boehler-Almax |
| Anvil Support Plate | Tungsten Carbide |
| Diamond Anvil Design | Boehler-Almax 3.30 mm – 85° |
| Pressure Mechanism | Gas Membrane |
| Maximum Pressure | > 100 GPa |
| Temperature Range | 300 K |
| Top Angle | X-Ray: conical 85º |
| Bottom Angle | X-Ray: conical 85º |
| Lateral Access | 2 x 130° azimuthal openings by 25° (axial) |
| Heating | N/A |
| Cooling | N/A |
| DAC Diameter | 50 mm |
| DAC Height | 34 mm |
| Working Distance to Sample | 15 mm |
| Numerical Aperture | 0.67 |
| Interfacing Solution | Via mounting pin (included) |
| DAC Weight | 300 g |
Maximum pressure is dependent upon the culet size of the anvils. Almax easyLab is committed to its policy of continuous improvement. Specifications may change without notice. easyLab and Diacell are registered trademarks of Almax easyLab.
FAQs
What is the maximum pressure I can achieve with the Bragg-(G) Plus and a given culet size?
| Culet size (µm) | 200 | 300 | 400 | 500 | 600 | 800 | 1000 |
| Bragg-(G) Plus Pmax (GPa) (*) | 114 | 98 | 63 | 41 | 28 | 16 | 10 |
(*) The Pmax values are only indicative. The maximum pressure achievable with a DAC is influenced by many others experimental parameters, like the gasket characteristics (material, thickness and hole size) or the pressure transmitting medium.
You may also like…
Share this product on:






















Reviews
There are no reviews yet.