Diamond anvils alignment

Good anvil alignment is needed for making reliable gaskets (proper pre-indentation and hole centring) and to ensure successful experiments. This page describes the steps to follow for diamond anvils alignment for a DAC.

Before starting the alignment

  • Check that the anvils are properly fixed in their supports;
  • Verify that all screws are correctly tightened;
  • Ensure both anvils are clean and free of debris (dust, metal particles, old gasket material).

Lateral alignment of the anvils

  • Bring the upper anvil close to the lower one, but do not put them in contact;
  • Observe both culets under the microscope;
  • Use the cell lateral screws to adjust the lower side anvil so the culets appear concentric;
  • The goal is to see two culets that overlap perfectly when viewed from above.

The video on the right shows an example on how to adjust the lateral alignment on a HeliosDAC diamond anvil cell.

Tilt alignment

  • Carefully bring the anvils in contact and observe the interference fringe pattern;
  • Note the point where the fringes are densest => one needs to adjust the anvil position in this area.
  • Note the DAC orientation, so as not to worsen the alignment with the adjustment (both X and Y directions are reversed under the metallurgical microscope);
  • Increase the distance between the anvils;
  • Use the DAC’s tilt screws to adjust the anvil position:
    • Tighten the screw (or screws) in the area where the fringes are densest. Tightening a screw pulls the anvil support up, thus increasing the diamonds separation. To be able to tighten the screw, one needs first to loosen the two opposite screws;
  • Repeat the above steps until only 3-4 large colored fringes are obtained.

Watch the video on the right for more tips on how to fine-tune the tilt alignment.

What good alignment looks like

  • Perfect overlap of culets;
  • No more than 3-4 wide colored fringes;
  • Uniform pre-indentation mark on both sides of the gasket;
  • Sample hole centered on both side of the gasket.

Signs of misalignment

  • Crescent shadows on one side;
  • Multitude of colored narrow fringes;
  • Off-center hole on one side of the gasket and centered on the other side;
  • Early anvils failure or gasket failure.